The
2019 ATS/ERS Spirometry Standards were recently released. The
standards are open-access
and can be downloaded
without charge from the October
15th
issue of the American Journal of Respiratory and Critical Care
Medicine. Supplements
are available from the same web
page.
The
2019 Spirometry Standards have been extensively re-organized with
numerous updates. Notably, a number of sections that were previously
discussed in the 2005 General Considerations for Lung Function
Testing have been updated and included in the 2019 Spirometry
Standards. Also notably, a number of stand-alone spirometry tests,
including the Flow-Volume Loop, PEF and MVV are not included in the
2019 Standards.
An
overview of changes and updates from the 2005 Spirometry Standards
are detailed within the 2019 Spirometry Standards (page e71, column
1, paragraph 2) and in the Data Supplement (pages E2-E3). In more
detail these include:
◆ The
list of indications for spirometry (page e73, table 1) was updated
primarily with changes in language.
-
“To
measure the effect of disease on pulmonary function” was updated
to “To measure the physiological effect of disease or disorder” -
“To
describe the course of diseases that affect lung function” was
updated to “To monitor disease progression” -
“To
monitor people exposed to injurious agents” was updated to “To
monitor people for adverse effects of exposure to injurious agents”
◆ Items
added to indications:
-
“Research
and clinical trials” -
“Preemployment
and lung health monitoring for at-risk occupations”
◆
Contraindications were
previously mentioned in the 2005 General Considerations rather than
the 2005 Spirometry Standards and these have been extensively updated
and expanded. Although the list of contraindications (page e74,
table 2) is fairly inclusive (and should be reviewed by all
concerned) there were items mentioned in the body of text that were
not in the table:
-
“Spirometry
should be discontinued if the patient experiences pain during the
maneuver.” -
“…because
spirometry requires the active participation of the patient,
inability to understand directions or unwillingness to follow the
directions of the operator will usually lead to submaximal test
results.”
◆ Notably,
abdominal aortic aneurysm (AAA) was not included as a
contraindication in the 2019 standards. (page e72, column 3,
paragraph 1)
◆ “Ambient
temperature, barometric pressure, and time of day must be recorded.”
(page e72, column 3, paragraph 2) This was mandated in the 2005
General Considerations and is now included in the 2019 Spirometry
Standards.
◆ Patient
testing considerations (page
e72, column 3, paragraph 3) have
been expanded and now includes:
-
“Testing
should preferably occur in a quiet and comfortable environment that
is separated
from the waiting room and other patients being tested.” -
“Drinking
water should be available. Tissues or paper towels should be offered
to help patients deal with secretions.” -
“A
smaller chair or a raised footstool should be provided for children
and small adults.”
◆ The
effects of testing in different positions is discussed
(page e73, column 1,
paragraph 1) and “If
testing is undertaken with the patient in another position, this must
be documented in the report.”
◆ Although mentioned in the 2005 General Considerations (pages 155-157) an abbreviated version of hygiene and infection control has been included in the 2019 Spirometry Standards (page e73, column 1, paragraph 1 and page e78, table 6).
◆ Spirometer
specifications must meet ISO 26782 standards. (page
e73, column 3, paragraph 3)
There are however, no significant differences in spirometer accuracy
(±
3%) from the 2005 Spirometry Standards. However
the 2005 Spirometry Standards noted that “if the [calibration]
syringe has an
accuracy of 0.5%, a reading of ±3.5%
is appropriate whereas in the 2019 Spirometry standards it is assumed
that it is a “±3%
accuracy tolerance, ±2.5%
for spirometers plus ±0.5%
for calibration syringes”.
◆ The
specifications for digitization of analog flow or volume signals are
higher than the 2005
Standards (“…sampling
rate must be ≥100
Hz with a minimum resolution of 12 bits.”, (page
e74, column 1, paragraph
1)).
This
sampling rate was indirectly specified in the proposed reporting
format in the 2005 Spirometry Standards (page 335, column
1, paragraph 2) but was
not specifically part of the spirometry equipment requirements.
◆ Manufacturers
are now required to provide an alert if a calibration is ±
2SD from the mean calibration factor or ±
6% from the previous calibration factor (page e75, column
1, paragraph 2).
◆ A
summary of mandated alert signals is included in the Data Supplement
(pages E30-E31).
◆ Quality
assurance now includes the statement “Precalibrated spirometers
cannot be recalibrated by the operator but must still undergo a
calibration verification. Manufacturers must specify the action to be
taken if a precalibrated device fails the calibration verification.”
(page e75, column 1, paragraph 1)
◆ Quality
assurance now includes the statement “Spirometry software must
include the ability to generate a report of calibrations that
includes the results of all verifications, the number of failed
calibration verifications in each session, and the changes in
calibration factors.” (page e75, column 1, paragraph 1)
◆ The 2019 Standard mandates that “The spirometry system must determine the
zero-flow level with the spirometer blocked before calibration,
calibration verifications, and patient tests.” (page e75, column
1, paragraph 3). Zero-flow levels were not discussed in the 2005
Standards.
◆ Quality
assurance now includes “Verification of reference value
calculations after software updates”. (page
e74,
table 3)
◆ Biological
QC is mentioned (page e75, column 2,paragraph 3) but it
is also indicated that “A biological control is not a substitute
for the use of a calibration syringe.” and that “In some
jurisdictions, including a biological control in quality control
reporting may constitute a breach of employee privacy protection.”
◆ Age
must now be reported in years to one decimal point. Height must now
be reported in centimeters to one decimal point. Weight must now be
reported in kilograms to the closest 0.5 kg. (page e75, column 3,
paragraph 2)
◆ Estimating
height from arm span or ulna length with reference equations is
discussed in the Data Supplement. (pages E27-E28).
◆ Patient
details now includes the statement “In persons aged 25 years or
older, for whom a reliable height measurement has been made
previously in the same facility, remeasuring height at subsequent
visits within 1 year may not be necessary.” (page e75, column 3,
paragraph 2)
◆ Patient details now includes a statement acknowledging transsexual / transgender patients: “When requesting birth sex data, patients should be given the opportunity to provide their gender identity as well and should be informed that although their gender identity is respected, it is birth sex and not gender that is the determinant of predicted lung size. Inaccurate entry of birth sex may lead to incorrect diagnosis and treatment.” (page e75, column 3, paragraph 3)
◆ In
the 2005 Standards the only smoking was indicated as an activity that
should be avoided prior to testing (page 327, column 3, paragraph 3).
The 2019 Standards indicates a number of additional activities that
should be avoided prior to testing (page e77, table 5).
◆ Like the 2005 Spirometry Standards, the 2019 Spirometry Standards notes “Well-fitting dentures are usually left in place.” but also notes that “…a larger 2018 study found that FVC was an average of 0.080 L higher when dentures were removed”
(page e76, column 2, paragraph 2)
◆ The
FVC maneuver now includes a fourth phase “inspiration at maximal
flow back to maximum lung volume” not present in the 2005
Standards. (page e76, column 3, paragraph 3)
◆ Expiratory
time is now either until EOT (End
of Test)
criteria has been met or when the expiratory reaches 15 seconds. The
2005 Spirometry Standard of “…and
the subject has tried to exhale for ≥3 s in children aged >10
yrs and for ≥6 s in subjects aged >10 yrs” no longer applies.
(page e77, column 1, paragraph 2)
◆ The
2019 Standard now mandates that “The
spirometry system must signal the operator when a plateau has been
reached or forced expiratory time (FET) reaches 15 seconds.” (page
e77, column 1, paragraph 2)
◆ The 2019 Spirometry Standard now indicates that “With appropriate coaching, children as young as 2.5 years old with normal cognitive and neuromotor function are able to perform acceptable spirometry” (page e77, column 2, paragraph 3) whereas the 2005 Standard stated “children as young as 5 yrs of age are often able to perform acceptable spirometry”.
◆ The back-extrapolation method has been updated to “At the point of PEF on the volume–time graph, a tangent is drawn with a slope equal to PEF, and its intersection on the abscissa defines Time 0, which becomes the start for all timed measurements.”
(page e77, column 3, paragraph 4)
◆ Extrapolated
Volume (EV) has been updated to Back Extrapolated Volume (BEV).
(page e77, column 3,
paragraph 4)
◆ The
minimum acceptable
level of back extrapolation has been updated to “<5%
of the FVC or 0.100 L, whichever is greater” from “<5%
of the FVC or 0.150 L, whichever is greater” (page
e77, column 3,
paragraph 4)
◆ Hesitation
time, which is not mentioned in the 2005 Standards and
is “…defined as the time from the point of maximal inspiration to
Time 0, should be 2 seconds or less” (page
e77, column 3, paragraph 4)
◆ End
of Test (EOT) acronym has
been replaced with End of Forced Exhalation (EOFE).
(page e78, column 2,
paragraph 1)
◆ EOFE
is considered acceptable when an expiratory time of 15 seconds has
been reached. Test equipment must acknowledge this expiratory time
with a double beep. (page e78, column 3, paragraph 2)
◆ EOFE is considered acceptable when “The patient cannot expire long enough to achieve a plateau (e.g., children with high elastic recoil or patients with restrictive lung disease). In this case, the measure of whether EOFE has been reached
is for the patient to repeatedly achieve the same FVC.” (page e78,
column 3, paragraph 3)
◆ The
2019 Spirometry standards include a discussion of the criteria
differences between acceptability and usability of spirometry
efforts. (page e79, table 7)
◆ The 2019 Spirometry Standards notes “Maneuvers that do not meet any of the EOFE acceptability criteria will not provide acceptable FVC measures. However, an acceptable FEV1 measurement may be obtained from a maneuver with early termination
after 1 second.” (page e79, column 1, paragraph 2)
◆ The 2019 Spirometry Standards includes the FIVC in the assessment of the FVC. Specifically “If the volume of the maximal inspiration (i.e., FIVC) after EOFE is greater than FVC, then the patient did not start the maneuver from TLC. FEV1 and FVC measurements from a maneuver with FIVC-FVC > 0.100 L or 5% of FVC, whichever
is greater, are not acceptable.” (page e79, column 3, paragraph 2)
◆ The
2019 Standards indicates that “The spirometry system software must
provide explicit feedback to the operator indicating FEV1 and FVC
acceptability at the completion of each maneuver.” (page
e80, column 2, paragraph 2) This
was suggested in the 2005 Standards but was not mandatory.
◆ The
2005 Standards included repeatability criteria
based on FVC volume (“For
those with an FVC of 1.0 L, both these values [the
difference between the largest and next largest FVC and FEV1] are
0.100 L (page 325, column
1, paragraph 7)). The
2019 standard uses this repeatability
criteria for ages ≤ 6 years, but not for any value of FVC. (page
e80, column 3, paragraph 2)
◆ Both
the 2005 and 2019 Spirometry Standards indicate that the maximum
number of FVC maneuvers in adults should be eight. The 2019
Standards, however, suggests that “When testing children, more than
eight attempts may be required because each attempt may not be a full
maneuver.” (page e81, column 1, paragraph 1)
◆ The
list of medications that should be withheld prior to reversibility
testing has been updated with significantly different withholding
times. (page e82, table 8).
◆ Recommended
reversibility testing protocols are detailed in the Data Supplement
(pages E32-E35).
◆ The
2005 Spirometry Standard discussed a variety of values that could be
reported whereas the 2019 Standard mandates those that must be
reported (page e82, table 9)
◆ The
2019 Spirometry Standard requires that FIVC be reported. (page e82,
table 9).
◆ The
selection and limitations of FEF25-75
is discussed in the 2019 Standards. (page e82, column 2, paragraph
2)
◆ The 2019 Standards recommends that the 2017 ATS reporting standards and the GLI reference equations be used for reporting spirometry results. (page e82, column 3, paragraph 2)
◆ The 2005 Spirometry Standards included a section on performing a flow-volume loop as a maneuver separate from spirometry (pages 326-328). The 2019 Standards mandates that the flow-volume loop is an integral part of spirometry. (page e82, column 3, paragraph 2)
◆ The
2005 Standards includes
a section on normal and abnormal flow-volume loops with 7 examples
(pages 327-328). This was not present
in the 2019 Standards. Flow-volume
loop examples are included in the Data Supplement (pages E4-E20).
◆ The
2019 Spirometry standards suggests but does not mandate that reports
should be exported as a .PDF file and that data should be exported
using the Clinical Document Architecture Release 2 standard of HL7
International or Fast Healthcare Interoperability Resources. (page
e82, column 3, paragraph 2)
◆ Database
elements are mandated and detailed in the Data Supplement (pages
E35-E38). The ability to export data in .XML format is mandated.
◆ The
date and time for each maneuver must be recorded ((page e82, column
3, paragraph 3)
◆ The
2019 Standard suggests that all reported spirometry efforts be graded
by repeatability.
(page e83, table 10) (page
e83, column 1, paragraph 2)
◆ The stability of tidal breathing prior to a slow VC maneuver is defined and its effect on IC is discussed in the 2019 Standards (“Stability is defined as having at least three tidal breaths with end-expiratory lung volume within 15% of the VT”). (page e84, column 1, paragraph 2). This was not previously discussed in the 2005 Standards.
◆ The
2019 Standards mandates that during a slow VC maneuver “The test
system must provide both a visual and an audible signal (single beep)
when a stable end-expiratory tidal lung volume is detected or there
have been 10 tidal breaths…” (page e83, column 1, paragraph 3)
◆ The 2019 Standards also mandates that during a slow VC maneuver the test system must provide “…for expiration to RV in either IVC or EVC maneuvers, a double beep when a plateau is reached (≤0.025 L in the last second) or the expiration
time reaches 15 seconds.” (page e83, column 2, paragraph 1)
◆ The
2005 Standards included the procedure for PEF testing (pages
330-331). This was not discussed in the 2019 Standards.
◆ The
2005 Standards included the procedure for MVV testing (page 331).
This was not discussed in the 2019 Standards.
*
* *
This
is a welcome update to the previous 2005 ATS/ERS Spirometry
Standards. Many deficiencies in the prior Standards have been
corrected and the new
standards are more pertinent to the current level of technology. As
usual, it will probably take a while before the existing software is
updated to reflect the new standards and in some cases this may never
occur.
I
was particularly pleased to see a discussion concerning
the acceptability and usability of spirometry efforts.
A spirometry effort does not have to be acceptable to be able to
provide clinically useful information.
Although
the selection criteria for reported results (largest FVC and largest
FEV1 from any acceptable
effort) has not changed
from the 2005 Standards (which is okay) there was no discussion on
how to indicate
that reported results
are a composite nor how to
link graphical results (flow-volume loops and volume-time curves) to
composite efforts nor how
to select other values (such as PEF) for a composite result.
I
would have liked to see some movement in the Standards towards
performing FEV1 and VC maneuvers separately. Interestingly
this was discussed very
briefly in the 2005
Standards (page 326, column 2, paragraph 3) but I could find no
similar discussion in the 2019 standards.
Although ethnicity was discussed in the Data Supplement (page E29) I would have liked to see a more in-depth discussion. Ethnicity was acknowledged by both the 2005 and 2019 Standards as being critical to the selection and use of reference equations but the guidelines are somewhat vague and many ethnicities are not represented in existing reference equations.
I
am slightly disappointed that the technical requirements for
spirometry systems has not changed from the 2005 Standards. This
however is probably realistic since
measuring
the flow and volume of exhaled air is technically quite difficult and
we’ve likely reached
a plateau in our ability to do this.
Despite
all this I am pleased
overall with the 2019 Standards. They are a distinct step forward
and should be welcomed by all.